The AvaSpec-256 Fiber Optic Spectrometer
Technical Data
Spectrometer platform
|
AvaSpec-256
|
AvaSpec-256-USB2
|
Optical Bench
|
Symmetrical Czerny-Turner, 45 mm focal length
|
Wavelength range
|
200-1100 nm
|
Resolution
|
0.4 -64 nm, depending on configuration (see table)
|
Stray light
|
< 0.2%
|
Sensitivity (AvaLight-HAL, 8 µm fiber) in counts/µW per ms integration time
|
30 (14-bit AD)
|
120 (16-bit AD)
|
Detector
|
CMOS linear array, 256 pixels
|
Signal/Noise
|
2000:1
|
AD converter
|
14 bit, 330 kHz
|
16 bit, 500 kHz
|
Integration time
|
1 msec - 60 seconds
|
0.6 msec - 10 minutes
|
Interface
|
USB 1.1, 12 Mbps
RS-232, 115.200 bps
|
USB 2.0 high speed, 480 Mbps
RS-232, 115.200 bps
|
Sample speed with on-board averaging
|
4 msec / scan
|
0.6 msec /scan
|
Data transfer speed
|
7-9 ms / scan (depending on # pixels transferred)
|
1.5 ms / scan
|
Digital IO
|
DB-15 connector, 2 Digital in, 12 Digital out
|
HD-26 connector, 2 Analog in, 2 Analog out, 3 Digital in, 12 Digital out, trigger, sync.
|
Power supply
|
12 VDC, reverse polarity protection, 160 mA (PS-12V/1.0A) or 5VDC USB power
|
Default USB power, 350 mA
Or with SPU2 external 12VDC, 350 mA
|
Dimensions, weight
|
175 x 110 x 44 mm (1 channel), 716 grams
175 x 165 x 85 mm (2 channel), 1.700 grams
|
175 x 110 x 44 mm(1 channel), 716 grams
|
Grating selection table for AvaSpec-256
|
Useable range
|
Spectral range (nm)
|
Lines/mm
|
Blaze (nm)
|
Order code
|
UV/VIS/NIR
|
200-1100*
|
900*
|
122
|
UV
|
UZ
|
UV/VIS/NIR
|
200-1100
|
400
|
300
|
300
|
UA
|
UV/VIS
|
200-850
|
200
|
600
|
300
|
UB
|
UV
|
200-750
|
100
|
1200
|
250
|
UC
|
UV/VIS
|
250-850
|
200
|
600
|
400
|
BB
|
VIS/NIR
|
300-1100*
|
800*
|
150
|
500
|
VZ
|
VIS/NIR
|
300-1100
|
400
|
300
|
500
|
VA
|
VIS
|
360-1000
|
200
|
600
|
500
|
VB
|
VIS
|
300-800
|
100
|
1200
|
500
|
VC
|
NIR
|
500-1050
|
200
|
600
|
750
|
NB
|
NIR
|
500-1050
|
100
|
1200
|
750
|
NC
|
NIR
|
600-1100
|
400
|
300
|
1000
|
IA
|
NIR
|
600-1100
|
200
|
600
|
1000
|
IB
|
* please note that not all 256 pixels will be used for the useable range
Resolution table (FWHM) for AvaSpec-256
Slit size (µm)
|
Grating (lines/mm)
|
25
|
50
|
100
|
200
|
500
|
122
|
5.5
|
7.9
|
15.7
|
32.0
|
79.0
|
150
|
4.5
|
6.4
|
12.8
|
26.0
|
64.0
|
300
|
2.5
|
3.2
|
6.4
|
13.0
|
32.0
|
600
|
1.0
|
1.5
|
3.0
|
6.0
|
16.0
|
1200
|
0.5
|
0.8
|
1.5
|
3.0
|
8.0
|
|