Detailed information for customers in Europe and America you can find at the manufacturer
|
|
|
| |
AvaSpec-NIR256 Fiber Optic Spectrometer
Technical Data
Spectrometer platform
|
AvaSpec-NIR256-1.7
|
AvaSpec-NIR256-2.2
|
AvaSpec-NIR256-2.5
|
Optical Bench
|
Symmetrical Czerny-Turner, 50 mm focal length
|
Wavelength range
|
900 - 1750nm
|
1000 - 2200nm
|
1000 - 2500nm
|
Resolution
|
6.0 - 50 nm
|
10.0 - 60 nm
|
15-90 nm
|
Stray light
|
< 0.1%
|
Sensitivity (AvaLight-HAL, 8 µm fiber) in counts (16-bit A/D)/µW per ms integration time
|
350
|
250
|
200
|
Detector
|
InGaAs linear array, 256 pixels
|
InGaAs linear array with 2 stage TE-cooling, 256 pixels
|
Signal/Noise
|
4000:1
|
1200:1
|
AD converter
|
16 bit, 500kHzHz
|
Integration time
|
0.52 ms
|
Interface
|
USB 2.0 high speed, 480 Mbps
RS-232, 115.200 bps
|
Sample speed with on-board averaging
|
1.06 msec /scan
|
Data transfer speed
|
1.56 msec /scan
|
Digital IO
|
HD-26 connector, 2 Analog in, 2 Analog out, 3 Digital in, 12 Digital out, trigger, synchronization
|
Power supply
|
Default USB power, 350 mA Or with SPU2 external 12VDC, 350 mA
|
External 100-240 VAC
|
Dimensions
|
175 x 165 x 85 mm
|
9.5" desktop
|
Grating selection table for AvaSpec-NIR256
|
Useable range
|
Spectral range (nm)
|
Lines/mm
|
Blaze (nm)
|
Order code
|
NIR
|
900-1750
|
750
|
200
|
1500
|
NIRA
|
NIR
|
1000-2200
|
980
|
150
|
2000
|
NIRZ
|
NIR
|
1000-2500
|
1470
|
100
|
2500
|
NIRY
|
Resolution table (FWHM) for AvaSpec-NIR256
Slit size (µm)
|
Grating (lines/mm)
|
50
|
100
|
200
|
500
|
100
|
15
|
25
|
50
|
90
|
150
|
10
|
15
|
30
|
60
|
200
|
6
|
12
|
24
|
50
|
|
| |
|
|
|