AvaSpec-3648 Fiber Optic Spectrometer


Technical Data

Optical Bench

Symmetrical Czerny-Turner, 75 mm focal length

Wavelength range

200 - 1100nm

Resolution

0.025 -20 nm, depending on configuration (see table)

Stray light

< 0.1%

Sensitivity (AvaLight-HAL, 8 µm fiber)

14000 counts (16-bit AD)/µW per ms integration time

Detector

CCD linear array, 3648 pixels

Signal/Noise

350:1

AD converter

16 bit, 1 MHz

Integration time

10 µsec - 10 min.

Interface

USB 2.0 high speed, 480 Mbps

RS-232, 115.200 bps

Sample speed with on-board averaging

3.7 msec /scan

Data transfer speed

3.7 msec /scan

Digital IO

HD-26 connector, 2 Analog in, 2 Analog out, 3 Digital in, 12 Digital out, trigger, synchronization

Power supply

Default USB power, 350 mA

Or with SPU2 external 12VDC, 350 mA

Dimensions, weight

175 x 110 x 44 mm(1 channel), 716 grams


Grating selection table for AvaSpec-3648

Use

Useable range

Spectral range (nm)

Lines/mm

Blaze (nm)

Order code

UV/VIS/NIR

200-1100**

900**

300

300

UA

UV/VIS

200-850

520

600

300

UB

UV

200-750

250-220*

1200

250

UC

UV

200-650

165-145*

1800

UV

UD

UV

200-580

115-70*

2400

UV

UE

UV

220-400

70-45*

3600

UV

UF

UV/VIS

250-850

520

600

400

BB

VIS/NIR

300-1100**

800**

300

500

VA

VIS

360-1000

500

600

500

VB

VIS

300-800

250-200*

1200

500

VC

VIS

350-750

145-100*

1800

500

VD

VIS

350-640

75-50*

2400

VIS

VE

NIR

500-1050

500

600

750

NB

NIR

500-1050

220-150*

1200

750

NC

NIR

600-1100**

500**

300

1000

IA

NIR

600-1100

500

600

1000

IB


* depends on the starting wavelength of the grating; the higher the wavelength, the bigger the dispersion and the smaller the range to select.


** please note that not all 3648 pixels will be used for the useable range


Resolution table (FWHM) for AvaSpec-3648

Slit size (µm)

Grating (lines/mm)

10

25

50

100

200

500

300

0.5

1.4

2.4

4.3

8.0

20.0

600

0.27

0.7

1.2

2.1

4.1

10.0

1200

0.07-0.13*

0.2-0.3*

0.4-0.6*

0.7-1.0*

1.4-2.0*

3.3-4.8*

1800

0.05-0.08*

0.12-0.21*

0.2-0.36*

0.4-0.7*

0.7-1.4*

1.7-3.3*

2400

0.04-0.07*

0.08-0.15*

0.14-0.25*

0.3-0.5*

0.5-0.9*

1.2-2.2*

3600

0.025-0.04*

0.07-0.10*

0.11-0.16*

0.2-0.3*

0.4-0.6*

0.9-1.4*











*depends on the starting wavelength of the grating; the higher the wavelength, the bigger the dispersion and the better the resolution